Documentation
Specifications
Application
Pharmaceuticals, pure and ultrapure water
Characteristic
Hygienic 2-electrode conductivity sensor
Measuring range
k=0,1: 0,04µS/cm - 500µS/cm
Measuring principle
Conductive conductivity cell
Design
Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable
Material
Isolation: PEEK
Electrode: polished stainless steel 1.4435
Sealing: Isolast (FFKM) FDA conform
Dimension
Electrode diameter: 17 mm
(0.66 inch)
Electrode length: 54 mm
( 2.10 inch)
Process temperature
Max. 120°C (150°C/30 min)
max. 248°F (302°F/30 min.)
Process pressure
max. 12 bar at 20°C
max.174 psi at 68°F
Temperature sensor
Integrated Pt100 or Pt1000
Ex certification
ATEX
Connection
Process connection: Varivent, Neumo, Clamp
Cable: TOP68 connector or fixed cable
Ingress protection
IP68
Certificates, declarations and approvals
- calibration certification - surface roughness Ra<0,8 (0,4)- Material certification 3.1.B - EHEDG, pharma certificate of compliance