GM700 emission monitoring solution

GM700 emission monitoring solution

Information

Information

Efficient process analysis – even under difficult conditions

Ammonia slip measurement in DeNOx plants

Emission monitoring of hydrogen fluoride in aluminum smelters

HCI measurement in combustion plants

Device properties:

High selectivity due to high spectral resolution

Short response times

No calibration required

No moving parts: minimal wear and tear

No gas sampling or conditioning required

In-situ measurement directly in the process for nearly uncompromised measured values

Measuring probe or cross-duct types for very good adaptation to the requirements of your measuring task

High reliability during operation

Can also be used in harsh ambient conditions

Enables the detection of quick and short-term process fluctuations

Specifications

Measured variables
NH3, HF, HCl, CO, H2O
Device version
Cross-duct version Measuring probe version
Ambient temperature range
–40 °C ... +50 °C Depends on parameterization; temperature change max. ±10 °C/h
Measuring principle
Diode laser spectroscopy (TDLS)
Measured variables
NH3, HF, HCl, CO, H2O
Performance-tested measurands
HF
Device version
Cross-duct version Measuring probe version
Certified measuring ranges
HF: 0 ... 5 mg/m³ / 0 ... 25 mg/m³ Suitability-tested in the cross-duct type
Measuring range
NH3: 0 ... 25 ppm / 0 ... 4,000 ppm HCl: 0 ... 7.5 ppm / 0 ... 3,000 ppm HF: 0 ... 3 ppm / 0 ... 2,000 ppm CO: 0 ... 100 ppm / 0 ... 30,000 ppm H2O: 0 ... 25 Vol.-% / 0 ... 100 Vol.-% Measuring ranges refer to 1 m measuring path Measuring ranges depend on application and device version
Ambient temperature range
–40 °C ... +50 °C Depends on parameterization; temperature change max. ±10 °C/h
Ambient humidity
≤ 95 % Relative humidity; non-condensing
Degree of protection
IP65/IP66/IP67
Conformities
Only for HF Approved for plants requiring approval 2001/80/EC (13. BImSchV) 2000/76/EC (17. BImSchV) 27. BImSchV German Clean Air Regulations EN 15267 EN 14181
Product safety
CE