Accuracy
Standard 0.2%
Optional 0.1%
Process temperature
-10°C...85°C
(14°F…212°F)
Pressure measuring range
100 mbar...10 bar
(1.5 psi...150 psi)
Process pressure / max. overpressure limit
40 bar (600 psi)
Main wetted parts
Alloy C
316L
optional coating AuPt
Max. measurement distance
100m (328ft) H2O
Material process membrane
316L, AlloyC
Gold-Rhodium
Measuring cell
100 mbar...10 bar
(1.5 psi...150 psi)
Measuring principle
Hydrostatic pressure
Characteristic
Pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences
Rod version
Supply voltage
4...20 mA HART
10,5...45 VDC (Non Ex):
Ex ia: 10,5...30 VDC
PROFIBUS PA:
9...32 VDC (Non Ex)
FOUNDATION Fieldbus:
9...32 VDC (Non Ex)
Reference Accuracy
Standard 0.2 %
Optional 0.1 %
Long term stability
0.05 % of URL/year
Process temperature
-10°C...85°C
(14°F…212°F)
Ambient temperature
-40°C...85°C
(-40°F...185°F)
Measuring cell
100 mbar...10 bar
(1.5 psi...150 psi)
Smallest calibratable span
10 mbar (1.45 psi)
Vacuum resistance
0 mbar abs.
Max. overpressure limit
40 bar (600 psi)
Process connection
Thread:
G1 1/2, MNPT1 1/2
Flange:
DN40...DN100,
ASME 2"...4",
JIS 10K
Material process membrane
316L, AlloyC
Gold-Rhodium
Material gasket
Viton, EPDM, Kalrez, none
Fill fluid
Inert oil,
Synthetic oil
Material housing
316L, Die-cast aluminum
Communication
4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
Safety approvals
Overfill protection WHG
SIL
Design approvals
EN10204-3.1
NACE MR0175
Drinking water approvals
KTW/ NSF
Specialities
Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions
Measuring principle
Hydrostatic
Characteristic / Application
Pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant and climatic-proofed
Rod version
Specialities
Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
diagnostic functions
Hermetically sealed Contite measuring cell (condensate-resistant and climatic-proofed)
Supply / Communication
4...20mA HART:
10,5...45V DC
Ex ia: 10,5...30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9...32V DC (Non Ex)
Accuracy
Standard 0.2%
Optional 0.1%
Long term stability
0.05 % of URL/ year
Ambient temperature
-40°C...85°C
(-40°F...185°F)
Process temperature
-10°C...85°C
(14°F…212°F)
Process pressure / max. overpressure limit
40 bar (600 psi)
Pressure measuring range
100 mbar...10 bar
(1.5 psi...150 psi)
Main wetted parts
Alloy C
316L
optional coating AuPt
Process connection
Thread:
G1 1/2, MNPT1 1/2
Flange:
DN40...DN100,
ASME 2"...4",
JIS 10K
Max. measurement distance
100m (328ft) H2O
Communication
4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI
Safety approvals
Overfill protection WHG
SIL
Design approvals
EN 10204-3.1
NACE MR0175
Drinking water approvals
KTW/ NSF
Options
Separate housing
Initial device settings
Overvoltage protection
Application limits
If pressurized, usage of two pressure transmitters to measure the differential pressure (electronic dP)
Observe ratio head pressure : hydrostatic pressure