Non-invasive RTD/TC thermometer with high measurement performance for demanding applications

Version with aluminium terminal head TA30A, 20° angle, DN25, 1 ", outside pipe ⌀ 33.7 mm

Mounting on the outer pipe wall: elimination of leakages, process contamination, flow disturbance

Version with explosion-proof terminal head TA30H, 90° angle, DN150, 6 ", outside pipe ⌀ 168.3 mm

Version with explosion-proof terminal head TA30H, 20° angle, DN25, 1 ", outside pipe ⌀ 33.7 mm

Version with stainless steel terminal head TA30R, 20° angle, DN25, 1 ", outside pipe ⌀ 33.7 mm

Version with cable probe TST310, 20° angle, DN25, 1 ", outside pipe ⌀ 33.7 mm

Thermal coupling element with advanced geometry and heat transfer material for optimized heat transfer; stainless steel clamps for easy installation

Ideal for energy monitoring

Can be used universally in all industries

Ideal for demanding process conditions such as high flow velocities, high process pressures, highly viscous or corrosive media, abrasion, pigging or small pipe diameters

Perfect for subsequent installation for measurements in existing facilities for energy and safety monitoring

Measurement accuracy and response time comparable with invasive measurements

No process opening required, no leakage risk

Increased safety for personnel, plant and the environment

User-friendly from product selection to installation and maintenance

Significant cost savings: shorter development and project planning times, reduced expenses for installation, certification and inspections as well as no costs for thermowells, nozzles and flanges, weld seam inspections and pipe extensions.

iTEMP temperature transmitter with all standard communication protocols and optional Bluetooth® connectivity

International certification: e.g. explosion protection according to ATEX, IECEx, CSA and NEPSI; functional safety (SIL)

Explore our unique product design and components

Comparison with invasive thermometers: accuracy and response time

Benefits of iTHERM SurfaceLine TM611

Specifications

Accuracy
depending on application Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
depending on application
Operating temperature range
-196 °C…400 °C (-320,8 °F...752 °F)
Measuring principle
Resistance Temperature Detector
Characteristic / Application
non-invasive thermometer cable probe or modular thermometer suitable for high process pressures surface mounting
Thermowell / protection tube
without (not intended to use with thermowell)
Insert / probe
non exchangeable RTD or TC insert
Outer diameter protection tube / Insert
Geometry of coupling element is designed according to diameter of process piping
Process connection
surface mounting
Operating temperature range
-196 °C…400 °C (-320,8 °F...752 °F)
Accuracy
depending on application Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
depending on application
Integration head transmitter
yes (4 … 20 mA; HART; iO-Link, PROFINET, PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
ATEX IECEx Ex tb, Ex ia, Ex db CSA C/US IS, NI UKCA Ex tb, Ex ia, Ex db NEPSI Ex ia, Ex d KC Ex d, Ex ia, Ex tb
Certification
SIL, MID, Marine