Vibronic Point level detection Liquiphant FTL80

Vibronic Point level detection Liquiphant FTL80

Category:Level
Information

Information

Point level switch for liquids for FailSafe overfill protection

Point level switch for liquids up to SIL3.

Process connections: Threads and flanges

Temperature: -60 to +280°C (-76 to +540°F)

Pressure: -1 to +100bar (-14.5 to +1,450psi)

Sensor material: 316L, Alloy C

International explosion protection certificates, overfill protection WHG, SIL, marine approvals

Use in safety systems requiring functional safety to SIL3 in accordance with IEC 61508 Ed.2.0/ IEC 61511-1/ISA 84-1 and DIN EN ISO 13849

Proof test: proof testing interval up to 12 years, Slave devices tested at the press of a button

Permanent self-monitoring/internal redundancy

4-20mA interface (acc. to NAMUR NE 06/NE 43): easy integration via the switching unit (Nivotester FailSafe FTL825) with a two-channel output (safety contacts) and locking function or direct integration into a safety PLC

No adjustment: quick, low-cost startup

No mechanically moving parts: no maintenance, no wear, long operating life

Monitoring of fork for damage, corrosion, build-up and mechanical blocking

Vibronic measuring principle

Functional safety SIL in process instrumentation

Liquiphant with permanent function check is the safe choice for point level detection in liquids

Specifications

Process temperature
-60°C...+280°C (-76°F...+540°F)
Process pressure / max. overpressure limit
Vacuum...100 bar (1450 psi)
Min. density of medium
Density from 0.4 g/cm3 (0.4 SGU)
Measuring principle
Vibration Liquids
Characteristic / Application
Compact point level switch for MIN and MAX safety applications up to SIL3, Dual Sealing/ Second line of defense Modular housing concept
Specialities
Outstanding features: - permanent LIVE signal monitors function safety - failsafe design and - high availability of the measuring values
Supply / Communication
2-wire 4...20 mA
Ambient temperature
-60°C...+70°C (-76°F...+160°F)
Process temperature
-60°C...+280°C (-76°F...+540°F)
Process pressure / max. overpressure limit
Vacuum...100 bar (1450 psi)
Min. density of medium
Density from 0.4 g/cm3 (0.4 SGU)
Main wetted parts
316L, Alloy C
Process connection
Flange: DN25...DN100 ASME 1"...4" JIS 10K...20K Thread: G3/4, G1, R3/4, R1, MNPT3/4, MNPT1
Sensor length
Compact
Communication
2-wire 4...20 mA
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, NEPSI, EAC
Safety approvals
Overfill protection WHG Leakage SIL3
Design approvals
EN 10204-3.1 NACE MR0175, MR0103 ASME B31.3 AD2000
Marine approval
GL/ ABS/ LR/ BV/ DNV
Options
PWIS free Pressure tight feed through Weather protection
Components
Nivotester FTL825, Transmitter