Memosens 2.0 contacting conductivity sensor for hygienic applications in life sciences and food

Memosens CLS16E measures conductivity in pure and ultrapure water applications for the monitoring and control of:

Water for injection (WFI)

Electro-deionizing

Ion exchangers

Reverse osmosis

Distillation

With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the digital sensor is suitable for hazardous area use.

Certified with quality certificate and EN 10204 3.1 and complying with EHEDG, USP Class VI, ASME and FDA, Memosens CLS16E is perfectly suited for hygienic applications in the life sciences and food industries.

The sensor is sterilizable and autoclavable, resists cleaning in place (CIP) and sterilization in place (SIP).

Stainless steel ensures robust, corrosion-free operation and the replaceable seal enables a long operating life.

Highest precision and measuring accuracy even at high temperatures and under pressure provide you with reliable data for optimum process and product quality.

Non-contact, inductive signal transmission ensures high process integrity.

IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.

The conductivity of a liquid can be measured using the conductive or toroidal measuring principles. This video shows what it is about and how these measuring principles work.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Документация

Спецификации

Measuring range
k=0,1: 0,04 to 500 µS/cm
Process temperature
-5 to 120 °C (23 to 248 °F) For Sterilizsation: max. 150 °C at 5 bar (Max. 302 °F at 73 psi)
Process pressure
13 bar at 20 °C (188 psi at 68 °F) 9 bar at 120 °C (130 psi at 248 °F)
Measuring principle
Conductive
Application
Pure and ultrapure water Pharmaceutical industry Final rinse WFI
Characteristic
Hygienic 2-electrode conductivity sensor
Measuring range
k=0,1: 0,04 to 500 µS/cm
Measuring principle
Conductive conductivity cell with electropolished stainless steel electrodes
Design
Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished
Material
n contact with medium Sensor depending on order execution: Electropolished, stainless steel 1.4435 (AISI 316L) PEEK Sealing depending on order execution: Form seal ISOLAST (FFKM) Form seal EPDM
Dimension
Electrode diameter: 17 mm (0.66 inch) Electrode length: 54 mm (2.13 inch)
Process temperature
-5 to 120 °C (23 to 248 °F) For Sterilizsation: max. 150 °C at 5 bar (Max. 302 °F at 73 psi)
Process pressure
13 bar at 20 °C (188 psi at 68 °F) 9 bar at 120 °C (130 psi at 248 °F)
Temperature sensor
Pt1000
Ex certification
ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
Connection
Process connection: Varivent, Neumo, BioControl, Clamp Sensor connection: Inductive, digital connection head with Memosens 2.0 technology
Ingress protection
IP68
Certificates, declarations and approvals
Inspection certificate 3.1, EN10204 Surface roughness test ISO4287/Ra (≤ 0,38 µm) Conformity to cGMP derived requirements ASME BPE US Food Contact Materials FDA CFR 21 EU Food Contact Materials (EG) 1935/2004 CN Food Contact Materials GB 4806 EHEDG EL Class I