RIA15 process indicator

RIA15 process indicator

Категория:System products
Информация

Информация

Loop-powered Ex ia process indicator as field or control panel device for 4 to 20 mA signals or HART®protocol

Display of 4 to 20 mA measured values or HART® process variables in all industries

Panel-mount housing or field housing as local display

Scalable display value

With the "Level" or "Analysis" options, basic settings can be made on selected sensors and available measured values and units can be preset quickly and easily.

Display of 4 to 20 mA measured values or optionally up to four of a sensor's HART® process variables in all industries; Use as primary or secondary HART® master

5-digit measured value display with 17 mm (0.67") character height with dimension and bargraph; backlight can be activated

Panel-mount housing (IP65) or field housing (IP66; Ex approval) as local display; Minimal installation depth

No external power supply required; Voltage drop ≤1 V (HART® ≤1.9 V)

Easy 3-key operation for configuring the device

International approvals; marine approval

SIL interference freeness acc. to EN 60508

Product video process display RIA15 with HART communication

Waterpilot FMX21 commissioning with RIA15 – easy and time saving

Explained in under 2 minutes - RIA15

Документация

Спецификации

Input
1 x analogue 4...20 mA/HART®
Output
Not defined
Display
LCD 17 mm (0.67") 5-digit 7-segment bargraph Plain text display for unit/TAG Activatable backlight
Power Supply
Loop powered
Measuring principle
Loop powered indicators
Measuring principle
Loop-powered field indicator
Function
Display
Display
LCD 17 mm (0.67") 5-digit 7-segment bargraph Plain text display for unit/TAG Activatable backlight
Power Supply
Loop powered
Input
1 x analogue 4...20 mA/HART®
Output
Not defined
Software functions
Not defined
Certificates
ATEX Ex ib [ia Ga] CSA C/US IS, NI CSA Ex ia FM IS EAC Ex ib IECEx Ex ib [ia Ga] JPN Ex ib NEPSI Ex ib UK Ex ib [ia Ga] ATEX IECEx Ex ib [ia Ga] marine certificate UL C/US IND Ex PESO Ex KC
SIL
Not defined
Panel mounting
field panel mounting