Документация
Спецификации
Characteristic / Application
metric style
modular temperature assembly
universal range of application
suitable for hazardous areas
flanged process connection
with neck
incl. thermowell / protection tube (metal)
usable with insert StrongSens
Thermowell / protection tube
welded protection tube
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
9,0 mm (0,35'')
11,0 mm (0,43'')
12,0 mm (0,47'')
14,0 mm (0,55")
15,0 mm (0,59")
Max. immersion length on request
up to 10.000,00 mm (393,70'')
Material protection tube/ thermowell
1.4404 (316L)
1.4571 (316Ti)
Alloy C276 (2.4819)
Alloy 600 (2.4816)
Optional coating
Tantal
PTFE Teflon
PVDF
Process connection
flange:
DN15 PN40 B1 (EN1092)
DN15 PN40 C (EN1092)
DN25 PN20 B1 (EN1092)
DN25 PN40 C (EN1092)
DN25 PN50 B1 (EN1092)
DN25 PN100 B2 (EN1092)
DN40 PN40 B1(EN1092)
DN50 PN40 B1 (EN1092)
ASME 1" 150 RF (B16.5)
ASME 1" 300 RF (B16.5)
Tip shape
straight
reduced
tapered
Surface roughness Ra
1,6 μm (63,0 μin.)
Operating temperature range
PT100 TF StrongSens:
-50 °C ...500 °C
(-58 °F ...932 °F)
PT100 WW:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
PT100 TF:
-50 °C ...400 °C
(-58 °F ...752 °F)
Max. process pressure (static)
at 20 °C: 50 bar (725 psi)
Accuracy
class A acc. to IEC 60751
class AA acc. to IEC 60751
Response time
depending on configuration
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
ATEX II1D Ex ia IIIC, II1G Ex ia IIC T6
ATEX II1/2D Ex ia IIIC,II1G Ex ia IIC T6
ATEX IECEx II1/2D Ex ta/tb IIIC Da/Db
ATEX II1G Ex ia IIC T6
ATEX II 3 G Ex nA IIC T6, II3D
NEPSI Ex ia IIC T6, Ex iaD 20 T85-T450
IECEx Ex ia IIC T6 Ga/Gb
EAC Ex ia IIC T6 Ga + DIP
NEPSI Ex nA IIC T6
Certification
Gost Metrology
SIL (transmitter only)