Analog conductivity sensor Condumax CLS15

Analog conductivity sensor Condumax CLS15

Kateqoriya:Liquid analysis
Məlumat

Məlumat

Conductive conductivity sensor for standard applications in pure and ultrapure water

Condumax CLS15 measures conductivity in pure and ultrapure water applications for the monitoring and control of:

Ion exchangers

Reverse osmosis

Cooling water

Distillation

Chip cleaning

The sensor can be used in power plants for:

Degassed conductivity

pH measurement with differential conductivity

With its international explosion protection approvals, such as ATEX and FM, the sensor is suitable for hazardous area use.

Reliable and accurate measuring values at low conductivities

Easy to clean thanks to polished measuring surfaces

Sterilizable

Robust design for high durability

Different cell constants provide a wide measuring range

Quality certificate stating the individual cell constant

The conductivity of a liquid can be measured using the conductive or toroidal measuring principles. This video shows what it is about and how these measuring principles work.

Spesifikasiyalar

Measuring range
k=0,01: 0-20 µS/cm k=0,1: 0-200 µS/cm
Process temperature
max. 130°C (150°C / 30 min.) (max. 266°F (302°F/30 min.)
Process pressure
max. 12 bar at 20°C (174 psi at 68°F)
Measuring principle
Conductive
Application
Water, power plant, process
Characteristic
2-electrode conductivity cell for pure and ultrapure water
Measuring range
k=0,01: 0-20 µS/cm k=0,1: 0-200 µS/cm
Measuring principle
Conductive conductivity cell with polished stainless steel electrodes
Design
2-electrode conductivity sensor with coaxially arranged electrodes, electropolished
Material
Electrode: 1.4435 Electrode shaft: PES
Dimension
Electrode diameter: 16 mm (0.62 inch) Electrode length: appr. 55 mm (2.14 inch)
Process temperature
max. 130°C (150°C / 30 min.) (max. 266°F (302°F/30 min.)
Process pressure
max. 12 bar at 20°C (174 psi at 68°F)
Temperature sensor
Pt100
Ex certification
ATEX
Connection
Process connection: 1/2" NPT, 3/4" NPT, 1 1/2" Clamp cable : 4-pole SXP connector or fixed cable
Ingress protection
IP67
Certificates, declarations and approvals
Calibration certification of the cell constante.