Analog conductivity sensor Condumax CLS16B

Analog conductivity sensor Condumax CLS16B

Kateqoriya:Liquid analysis
Məlumat

Məlumat

Contacting conductivity sensor for hygienic applications in the food & beverage and life sciences industries

Condumax CLS16B measures conductivity in pure and ultrapure water for the monitoring of:

Ion exchangers

Reverse osmosis

Distillation

Water for injection (WFI)

With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the analog sensor is suitable for hazardous area use.

Perfectly suited to life sciences and food & beverage: Complies with EHEDG, ASME, FDA and USP Class VI

Hygienic: Autoclavable, sterilizable and suitable for cleaning-in-place (CIP) and sterilization-in-place (SIP)

Accurate: Cell constant of the sensor is measured individually

Robust: Electropolished stainless steel body for corrosion-free operation and easy cleaning

Long-operating life: Replaceable, hygienic EPDM/PEEK sealing

The conductivity of a liquid can be measured using the contacting or inductive measuring principles. This video shows what it is about and how these measuring principles work.

Spesifikasiyalar

Measuring range
k=0,1: 0,04µS/cm - 500µS/cm
Process temperature
Max. 120°C (150°C/30 min) max. 248°F (302°F/30 min.)
Process pressure
max. 12 bar at 20°C max.174 psi at 68°F
Measuring principle
Conductive
Application
Pharmaceuticals, pure and ultrapure water
Characteristic
Hygienic 2-electrode conductivity sensor
Measuring range
k=0,1: 0,04µS/cm - 500µS/cm
Measuring principle
Conductive conductivity cell
Design
Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable
Material
In contact with medium Sensor depending on order execution: Electropolished, stainless steel 1.4435 (AISI 316L) PEEK Sealing depending on order execution: Form seal ISOLAST (FFKM) Form seal EPDM
Dimension
Electrode diameter: 17 mm (0.66 inch) Electrode length: 54 mm ( 2.10 inch)
Process temperature
Max. 120°C (150°C/30 min) max. 248°F (302°F/30 min.)
Process pressure
max. 12 bar at 20°C max.174 psi at 68°F
Temperature sensor
Integrated Pt100 or Pt1000
Ex certification
ATEX
Connection
Process connection: Varivent, Neumo, Clamp Cable: TOP68 connector or fixed cable
Ingress protection
IP68
Certificates, declarations and approvals
Inspection certificate 3.1, EN10204 Surface roughness test ISO4287/Ra (≤ 0,38 µm) Conformity to cGMP derived requirements ASME BPE US Food Contact Materials FDA CFR 21 EU Food Contact Materials (EG) 1935/2004 CN Food Contact Materials GB 4806 EHEDG EL Class I