Məlumat
Məlumat
Metric RTD thermometer with connecting cable for use in many process and laboratory applications
The resistance thermometer is specially suited to temperature measurement in machinery, laboratory equipment and plants in gaseous or liquid media like air, water, oil and others.
High flexibility through user-specific insertion lengths and variable process connections
Fast response time
Single or double Pt100 sensor of accuracy class A, B, or AA as per IEC 60751
Types of protection for use in hazardous locations: Intrinsic safety (Ex ia) and non-sparking (Ex nA)
Marine approval
Sənədlər
Spesifikasiyalar
Accuracy
class A acc. to IEC 60751
class B acc. to IEC 60751
class AA acc. to IEC 60751
Response time
t50 = 3 s
t90 = 6 s
Max. process pressure (static)
at 20°C: up to 75 bar (depending on the process connection)
Operating temperature range
PT 100:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
Max. immersion length on request
up to 10.000,00 mm (393,70'')
Measuring principle
Resistance Temperature Detector
Characteristic / Application
metric style
cable probe
suitable for hazardous areas
process connection as compression fitting
Thermowell / protection tube
without (not intended to use with thermowell)
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
3,0 mm (0,12'')
6,0 mm (0,24'')
Max. immersion length on request
up to 10.000,00 mm (393,70'')
Material protection tube/ thermowell
1.4404 (316L)
1.4571 (316Ti)
Process connection
male thread:
G1/4"
G1/2"
M10x1
M8x1
compression fitting:
G1/8"
G1/4"
G1/2"
NPT1/4"
NPT1/2"
M10x1
M8x1
Tip shape
straight
Operating temperature range
PT 100:
-200 °C ...600 °C
(-328 °F ...1.112 °F)
Max. process pressure (static)
at 20°C: up to 75 bar (depending on the process connection)
Accuracy
class A acc. to IEC 60751
class B acc. to IEC 60751
class AA acc. to IEC 60751
Response time
t50 = 3 s
t90 = 6 s
Integration head transmitter
no
Ex - approvals
ATEX Ex ia, Ex nA
IECEx Ex ia
NEPSI Ex ia
EAC Ex ia
NEPSI Ex ia
UKCA Ex ia, Ex nA, Ex tc