Memosens CCS50D

Memosens CCS50D

Kateqoriya:Liquid analysis

Spesifikasiyalar

Application
Adequate disinfection in drinking water, preventing of pathogen formation in cooling water systems, water used to wash pre-packaged vegetables, ensures the absence of chlorine dioxide in beverage systems
Characteristic
Amperometric measurement of dissolved chlorine dioxide
Measuring range
Trace: 0 to 5 mg/l ClO2 Standard: 0 to 20 mg/l ClO2 High: 0 to 200 mg/l ClO2
Measuring principle
Closed, membrane covered measuring cell Reduction of chlordioxide (ClO2) to chloride at the cathode with appr. 120mV
Design
Closed amperometric 2-electrode measuring cell with PVDF membrane
Material
Sensor shaft: POM Membrane: PVDF Membrane cap: PVDF
Dimension
Diameter: 25 mm (0.98 inch) Length: 161 mm (6.34 inch)
Process temperature
0 to 55 °C , non-freezing (32 to 130 °F)
Process pressure
Max. 2 bar abs (Max. 29 psi abs)
Temperature sensor
10k NTC integrated (Memosens)
Connection
Inductive, digital connection head with Memosens