S700 process gas analyzer

S700 process gas analyzer

Kateqoriya:Optical analysis
Məlumat

Məlumat

Tailor-made gas analysis for process and emission monitoring

Cooling gas monitoring of turbo generators

CO measurement for coal mill monitoring

H2 purity measurements in pressure swing adsorption systems

CO2 monitoring in natural gas during natural gas processing

Device properties:

Various analyzer modules are available: UNOR (NDIR), MULTOR (NDIR), OXOR-E (electrochemical O2), OXOR-P (paramagnetic O2), THERMOR (thermal conductivity)

Over 60 measuring components to choose from

Different housing variants

Up to three analyzer modules in one housing

Simple application-specific configuration thanks to modular design

Can also be used in Zone 1 and Zone 2 (ATEX) explosion-hazardous areas

Automatic readjustment using test gas or calibration cell

Integrated self-monitoring and error analysis

Process monitoring for the safety and quality of manufacturing processes and products by measuring flammable gases and explosive gas mixtures in zones 1 and 2

Reliable remote control and integration into distributed control systems thanks to optional equipment with various interface converters

Spesifikasiyalar

Measured variables
CH4, CO, CO2, H2, H2O, N2O, NH3, NO, NO2, O2, SO2 Hydrocarbons (e.g. C2H2), halogenated hydrocarbons (e.g. CH2Cl2) and other gases upon request
Ambient temperature range
+5 °C ... +45 °C
Hazardous area approvals
S715 Ex: ATEX II 3G Ex nR IIC T6 Gc Without intrinsically safe measured value outputs ATEX II 3G Ex nR [ib] IIC T6 Gc With intrinsically safe measured value outputs S720 Ex / S721 Ex: ATEX II 2G Ex db ia IIC T6 Gb Without intrinsically safe measured value outputs ATEX II 2G Ex db ia [ia] IIC T6 Gb With intrinsically safe measured value outputs
Measuring principle
NDIR spectroscopy, Electrochemical cell, Paramagnetic dumbbell principle, Thermal conductivity measurement
Measured variables
CH4, CO, CO2, H2, H2O, N2O, NH3, NO, NO2, O2, SO2 Hydrocarbons (e.g. C2H2), halogenated hydrocarbons (e.g. CH2Cl2) and other gases upon request
Performance-tested measurands
CO, NO, O2, SO2
Process pressure
Hosed internal gas paths:–20 kPa ... 30 kPa Piped internal gas paths:–20 kPa ... 100 kPa With “OXOR-E” analyzer module:–20 kPa ... 30 kPa S720 Ex/S721 Ex:–20 kPa ... 10 kPa
Ambient temperature range
+5 °C ... +45 °C
Ambient humidity
≤ 95 % Relative humidity; non-condensing
Conformities
Industrial Emissions Directive (IED) CCEP U.S. EPA conform China's EPA compliant
Hazardous area approvals
S715 Ex: ATEX II 3G Ex nR IIC T6 Gc Without intrinsically safe measured value outputs ATEX II 3G Ex nR [ib] IIC T6 Gc With intrinsically safe measured value outputs S720 Ex / S721 Ex: ATEX II 2G Ex db ia IIC T6 Gb Without intrinsically safe measured value outputs ATEX II 2G Ex db ia [ia] IIC T6 Gb With intrinsically safe measured value outputs
Degree of protection
IP65 / NEMA 7
Product safety
CE, CSA (option)